Automated Immunochemical Method for Determination of Urinary Protein of Plasma Origin
[摘要] An automated continuous-flow procedure has been developed for the rapid determination of urinary proteins of plasma origin. Antiserum to whole human plasma was used as the reagent, and the antigen—antibody reactions were quantitated by nephelometry. By adding polyethylene glycol (mol wt 6000-7500) to the reaction medium, reaction time was decreased to <3 min; no sample blanks were required; and samples were analyzed at a rate of 70 per hour. Recovery studies yielded an average of 98.5% of the added protein. In-run replicate precision (CV) of the method was 1.45%; day-to-day precision was 2.58%.