Demonstration of the Pre–Emergency Use Authorization Path Using 3 Minor Groove Binder–Hydrolysis Probe Assays to Detect Escherichia coli O104:H4
[摘要] BACKGROUND: The Department of Defense (DoD) and the Food and Drug Administration (FDA) have collaboratively worked on a pre–Emergency Use Authorization (pre-EUA) process for in vitro diagnostic (IVD) devices, using FDA's regulatory flexibilities under the EUA authorities. The pre-EUA process enables FDA review of data in anticipation of a request for an EUA, advancing US government public health emergency preparedness efforts.METHODS: The IVD device developed to detect Escherichia coli O104:H4, for which an EUA has not been issued, serves as an example to illustrate that process. Specifically, DoD designed real-time PCR assays to target the virulent E. coli strain O104:H4 (etiological agent of the 2011 German outbreak) including: fliC (flagellin), Agg3C ( AAF ), and rfb ( wbwC ) on the basis of the published sequences.RESULTS: After development and optimization of these 3 specific assays, a defined protocol was followed to determine and document the sensitivity and specificity of each assay analytically.CONCLUSIONS: FDA reviewed these data and returned commentary on additional required experiments to complete the pre-EUA process and expedite the use of the device should there be an emergency need for an IVD device to detect this virulent E. coli strain before such a test is cleared by FDA.