Failure of Processed Quality-Control Materials to Detect Interference from Labile Form of Hemoglobin A1c
[摘要] Our laboratory uses the Bio-Rad VariantTM system to measure hemoglobin A1c. The procedure begins with an autodiluter aspirating 5 μL of blood and dispensing it with 1 mL of hemolyzing reagent. The hemolyzing reagent also contains chemicals to remove the Schiff base-labile form (pre-A1c) (1).A technologist incorrectly connected wash solution to the autodiluter. Bio-Rad’s hemolysis reagent, wash solution, and buffer solutions all come in rectangular clear plastic bottles of identical shape and size. The bottles are labeled on only one side, and that side of the hemolysis reagent bottle lies next to the autodiluter (to minimize the distance the plastic tubing runs between the front of the autodiluter and the bottle cap, allowing the tubing to reach the bottom of the bottle). Therefore, no other technologists saw that …