Toward absolute methods in clinical chemistry: application of mass fragmentography to high-accuracy analyses.
[摘要] We review recently developed reference methods based on mass fragmentography (specific ion monitoring) with use of isotope-labeled internal standard, for determination of cholesterol, triglycerides, urea, glucose, cortisol, progesterone, and testosterone. With an optimal ratio between standard and material to be determined, the relative standard deviation of these methods, based on duplicate determinations, is between 1.3 and 2.7%. The very high specificity of the methods, in combination with the fact that the ratio between labeled and unlabeled molecules is determined with a high degree of accuracy, makes it likely that the most significant errors in the methods are related to errors in pipetting. The possibility is discussed that these methods might be developed into "absolute" or "definitive" methods by subjecting each individual step in the determination to a detailed analysis with respect to possible error. We also discuss some of the consequences of the different reference methods on the choice of routine methods.