Unsuitability of porcelain crucibles for dry ashing of biological tissues for calcium analysis.
[摘要] At temperatures required for complete release of calcium from beef liver by dry ashing, porcelain crucibles release significant amounts of calcium into the ash, which leads to erroneously high calcium values in the samples. Heated, empty porcelain crucibles do not show released calcium. Both fused silica and platinum crucibles fail to show any significant interaction with the sample. Samples dry-ashed in fused silica crucibles were compared to samples subjected to the hot trichloracetic acid-acetic acid extraction method, which gave results not significantly different if the tissue was first minced. However, calcium was incompletely extracted by the hot acid from whole segments of liver weighing about 0.2 g.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]