已收录 268921 条政策
 政策提纲
  • 暂无提纲
Ultrasensitive Flow-based Immunoassays Using Single-Molecule Counting
[摘要] Background: Immunoassay (IA) technology has expanded the clinical utility of protein biomarkers, but demands for increased sensitivity, dynamic reporting ranges, and small sample volumes have limited the potential clinical usefulness of many biomarkers. We assessed the performance, including limits of detection (LODs) and the dynamic reporting range, of an IA-based technology, Erenna Immunoassay System, for a series of biomarkers, including cardiac troponin I (cTnI).Methods: Erenna IAs were used with 10 different and clinically important biomarkers to ascertain the LOD with various sample sizes (10 μL to 200 μL).Results: The Erenna Immunoassay System generated LODs of 10–100 pg/L using 100 μL of sample. For cTnI, the LOD was 0.2 ng/L and a 10% CV was seen between 0.78 and 1.6 ng/L.Conclusions: The Erenna IA-based technology reproducibly measures protein biomarkers with detection limits of 10–100 pg/L, with a dynamic range of >4.5 logs in sample volumes of 50–150 μL.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文