已收录 268921 条政策
 政策提纲
  • 暂无提纲
Optimized assay for serum angiotensin-converting enzyme activity.
[摘要] This optimized two-point kinetic assay for serum angiotensin-converting enzyme is based on the colorimetric determination of hippurate with cyanuric chloride/dioxan reagent. The hippurate is released from hippuryl-L-histidyl-L-leucine by angiotensin-converting enzyme in the presence of chloride ion. CVs for the method (within-run and between-run) ranged from 2.1 to 3.2%. Linearity extends up to 200 U/L. Results are unaffected by lipemia and icterus. Hemoglobin in concentrations greater than 1.5 g/L causes a slight negative interference. Reference intervals for men and women are 22-82 U/L and 25-69 U/L, respectively.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文