已收录 268921 条政策
 政策提纲
  • 暂无提纲
A candidate reference method for determination of total protein in serum. II. Test for transferability.
[摘要] The transferability of the candidate Reference Method for total serum protein was tested in eight laboratories in the United States and Europe. National Bureau of Standards SRM 927 (bovine serum albumin) was used in each analytical run as the calibration standard. The mean absorptivity value obtained for this material was 0.2983 L g-1 cm-1. Four serum pools prepared at the Centers for Disease Control were analyzed on each of 15 days. Within-run variation of the protein values (expressed as CV) in the eight laboratories ranged from 0.1 to 2.5% and day-to-day (total) variation in six of the laboratories ranged from 0.4 to 1%.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文