Clinical Evaluation of a Modified "Oxford T4-by-Column" Method for Serum Thyroxine
[摘要] Three commercially available kits for determining serum thyroxine (T4) by column were evaluated. Results with all were less than satisfactory, but a series of changes in the Oxford method resulted in increased reproducibility, substantially improved recovery, and extended linearity of the standard curve. Results compared well with the thyrometabolic status of patients and normal individuals. The 95% confidence limits for 219 "normal" hospital employees were 2.9-6.2 µg/100 ml. The suggested normal range for 241 clinically euthyroid patients is 2.5-6.0 µg/100 ml. The coefficient of correlation between results of the proposed method and of the PBI method was 0.30; for the T4 by the Murphy—Pattee method it was 0.75.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]