已收录 268920 条政策
 政策提纲
  • 暂无提纲
Manual and Automated Determination of Immunoglobulins in Unconcentrated Cerebrospinal Fluid
[摘要] Manual procedures have been developed for determination in human cerebrospinal fluid of the immunoglobulins IgG, IgA, and IgM. An automated continuous-flow procedure for determination of IgG in human CSF is also described. The immunoglobulins are quantitated in both cases by immunoprecipitin reactions, with use of commercial antiserum. Values determined for the immunoglobulins by the manual method correlate with values for gamma globulin obtained by electrophoresis. The values for IgG from the manual method also correlate with IgG values determined by the automated procedure. The precision of duplicate determinations with the manual procedures for IgG, IgA, and IgM is 3.4, 1.8, and 1.0% (relative standard deviation), respectively, and that for lgG with the automated system is 3.48%.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文