已收录 272918 条政策
 政策提纲
  • 暂无提纲
An Improved Automated Determination of Serum Total Cholesterol with a Single Color Reagent
[摘要] An anhydrous color reagent containing ferric chloride is pumped through Tygon tubing to a glass coil in a 95° heating bath. Manually prepared serum extracts (1:20 isopropanol dilution) are presented to the stream of preheated color reagent, and the two are passed through three mixing coils, connected in series. The absorbance of the resultant color is determined at 550 mµ in a 15 mm. tubular flow-cell.The improved N-automated procedure gave values of greater precision (5.6%) for total cholesterol determined in replicate samples from sera pools than the N-automated procedure which lacked precision (27.7%). The values found in 60 individual serums are within 6.0% of total cholesterol values determined by the Abell et al. method.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文