已收录 268921 条政策
 政策提纲
  • 暂无提纲
Evaluation of an Enzymatic Procedure for Determination of Serum Cholesterol with the Abbott ABA-100
[摘要] A previously reported direct enzymatic method for cholesterol [ Clin. Chem. 20, 470 (1974)] was critically evaluated in the clinical laboratory. Other methods compared to this enzymatic procedure were: ( a ) the manual Abell—Kendall procedure, giving a regression slope (proportional sensitivity) of 0.998 and a negative bias of 9.7 mg/dl in the enzymatic procedure; ( b ) Lipid Research Clinic procedure, giving a regression slope of 1.005 and a negative bias of 8.8 mg/dl in the enzymatic procedure; and ( c ) the SMA 12/60 (Technicon) method, giving a regression slope of 0.881 and a negative bias of 20.8 in the enzymatic procedure. Replicate analyses of pooled serum by the enzymatic procedure demonstrated the following inter-run precision: mean = 178 mg/dl, SD = 4 mg/dl, CV = 2.2%. Bilirubin caused a slight positive interference (1 mg of bilirubin per deciliter appears as 1-2 mg of cholesterol). Hemolysis, lipemia, and a series of steroid drugs caused no interference. A group of 465 normal subjects was tested by the enzymatic procedure. This population had a mean cholesterol of 203 mg/dl. The 2.5 percentile was 130 mg/dl, and the 97.5 percentile was 316 mg/dl.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文