已收录 268921 条政策
 政策提纲
  • 暂无提纲
A coaxial extractive electrospray ionization source
[摘要] An extractive electrospray ionization (EESI) source design is presented to improve the reproducibility of the interactions occurring in EESI. This design uses three concentric capillaries to deliver the solvent, sample and nebulizing gas. Coaxial EESI was found to improve the inter-experiment variation by a factor of four and intra-run relative standard deviation by a factor of 2.4. Unlike most standard EESI designs, the device has the form factor of a standard electrospray emitter and can be used without any further instrument modifications.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文