Rare complications of osteolysis and periprosthetic tissue reactions after hybrid and non-hybrid total disc replacement
[摘要] PurposeFew complications have been reported for lumbar total disc replacement (TDR) and hybrid TDR fixations. This study evaluated retrieved implants and periprosthetic tissue reactions for two cases of osteolysis following disc arthroplasty with ProDisc-L prostheses.MethodsImplants were examined for wear and surface damage, and tissues for inflammation, polyethylene wear debris (polarized light microscopy) and metal debris (energy-dispersive X-ray spectroscopy).ResultsDespite initial good surgical outcomes, osteolytic cysts were noted in both patients at vertebrae adjacent to the implants. For the hybrid TDR case, heterotopic ossification and tissue necrosis due to wear-induced inflammation were observed. In contrast, the non-hybrid implant showed signs of abrasion and impingement, and inflammation was observed in tissue regions with metal and polyethylene wear debris.ConclusionsIn both cases, wear debris and inflammation may have contributed to osteolysis. Surgeons using ProDisc prostheses should be aware of these rare complications...
[发布日期] [发布机构]
[效力级别] [学科分类] 骨科学
[关键词] Chronic inflammation ;Hybrid fixation ;Metal wear debris ;Osteolysis ;Polyethylene wear debris ;Total disc replacement [时效性]