Evaluation of impingement behaviour in lumbar spinal disc arthroplasty
[摘要] IntroductionThe objective of our in vitro study was to introduce a test method to evaluate impingement in lumbar spinal disc arthroplasty in terms of wear, contact pattern, metal ion concentration and particle release.Material and MethodImpingement wear simulation was performed on a 6-station spinal wear simulator (Endolab, Germany) on a lumbar spinal disc system (activ® L Aesculap AG, Germany) using four different protocols specific to impingement in flexion, in extension, in lateral bending and in combined flexion bending. Impingement contact stress is intentionally created by applying an angular displacement of +2° in addition to the intended range of motion in the impingement direction, whereas a bending moment of 8 Nm remains constant during the impingement phase (plateau).ResultsAn average volumetric wear rate of 0.67 mm3/million cycles was measured by impingement under flexion, of 0.21 mm3/million cycles under extension, of 0.06 mm3/million cycles under lateral bending and of 1.44 mm3/million cycles under combined flexion bending. The particle size distribution of the cobalt-chromium wear particles released by impingement in flexion (anterior), extension (posterior), lateral bending (lateral) and combined flexion bending (antero-lateral) revealed that most of the detected cobalt-chromium particles were in a size range between 0.2 and 2 µm.ConclusionThe impingement wear simulation introduced here proved to be suitable to predict in vivo impingement behaviour in regard to contact pattern seen on retrieved devices of the activ® L lumbar disc arthroplasty design in a pre-clinical test...
[发布日期] [发布机构]
[效力级别] [学科分类] 骨科学
[关键词] Lumbar total disc arthroplasty ;Impingement ;Wear simulation ;Pattern of contact ;Particle analysis [时效性]