Reproducibility of the cervical range of motion (CROM) device for individuals with sub-acute whiplash associated disorders
[摘要] PurposeThe objective of this study was to evaluate the reproducibility of the cervical range of motion device when measuring both active and passive range of motion in a group of individuals with sub-acute Whiplash Associated Disorders (WAD).MethodsParticipants were recruited as part of a large multi-centre Randomised Controlled Trial from UK emergency departments. Experienced research physiotherapists measured active and passive cervical spine movements in all directions. Both intra- and inter-observer reliability and agreement were assessed using the intra-class correlation coefficient, standard error of measurement and limits of agreement methods.ResultsDifferent groups of 39 and 19 subjects were included in the intra and inter-observer studies, respectively. The CROM device demonstrated substantial intra- and inter-observer reliability and agreement for all the active and passive half-cycle movements (ICC range 0.82–0.99) with the exception of one (passive right lateral flexion for inter-observer; ICC 0.77).ConclusionsThe CROM device has proven to be a reproducible measurement method for a symptomatic WAD population using the measurement protocol described and can be used with confidence to differentiate individuals according to a single measurement...
[发布日期] [发布机构]
[效力级别] [学科分类] 骨科学
[关键词] MeSH term ‘Reproducibility of results’ ;Cervical spine ;CROM [时效性]