已收录 268921 条政策
 政策提纲
  • 暂无提纲
Rapid and automated sequential determination of ultra-trace long-lived actinides in air filters by inductively coupled plasma mass spectrometry
[摘要] A novel, rapid and completely automated method for the determination of long-lived actinides in air particulate samples was developed using ion chromatography separation prior to ICP-MS analyses. U, Pu and Am were pre-concentrated and separated from radioactive and stable interferences through a sequential arrangement of three columns (TEVA, U/TEVA, DGA) arranged in order to allow selective retention of the radioisotope on the designated resin. Detection and quantification of the various radioisotopes were performed on-line using inductively coupled plasma mass spectrometry (ICP-MS). Detection limits of 0.0006 (238U), 0.0063 (239Pu), 0.0041 (240Pu), 0.062 (241Am) μBq/m3 were obtained for air filters (based on 3000 m3 of air filtered). The automated procedure has been successfully tested on SRMs and spiked samples and showed high recovery yields (>90%). A total analysis time of 18 min is required for the separation, detection, and column rinsing/regeneration for subsequent analysis...
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文