已收录 272983 条政策
 政策提纲
  • 暂无提纲
Double-codified nanogold particles based automated flow-through CLEIA for 2,4-dinitrotoluene
[摘要] A sensitive, automated micro flow-through chemiluminescence enzyme immunoassay (CLEIA) was developed for the determination of 2,4-dinitrotoluene (DNT) which employed a DNT analog covalently immobilized on a glass microchip, and nanogold particles (AuNPs) modified with HRP-labeled anti-mouse antibodies. The HRP-luminol-H2O2 chemiluminescent system (because of its high sensitivity) was chosen as the detection system. The obtained chemiluminescence (CL) intensity was inversely proportional to the DNT concentration in the sample. The working range for DNT quantification was 0.01–1 μg mL−1. Both the intra- and inter-assay coefficients of variation were less than 10%. The proposed assay had a LOD of 0.009 μg mL−1, which was 3-fold lower than a traditional HRP-labeled anti-mouse antibody-based CLEIA. The overall assay time was 13 min for each sample, enabling relatively rapid analysis...
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文