不同预处理方法对复用管腔器械清洗效果的比较
[摘要] Abstract:Objective To compare the cleaning effect of two different pretreatment methods on reused lumen instruments. Methods 〓Totally 638 lumen instruments of surgical aspirator and intrauterine aspirator were randomly divided into routine group and experimental group. In the routine group, 319 instruments were scrubbed and automated cleaning after soaking with enzymatic cleaner. In the experimental group, 319 instruments were scrubbed and automatic cleaning after soaking with alkaline cleaner. The cleaning effect was detected by long swab test and ATP bioluminescence method. Results 〓The result of visual inspection showed that the qualified rate of cleaning method in experimental group was higher than that in routine group. The ATP bioluminescence test results of the instruments met the standard after cleaning two methods, and the qualified rate of RLU value in the experimental group was obviously superior to that of the conventional group. Conclusion 〓The method of scrub and automated cleaning after soaking with alkaline cleaner is better in cleaning of reused lumen instruments.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 护理学
[关键词] 复用管腔器械;预处理;清洗效果;消毒 [时效性] 