Case Report of Multiembolic Cerebrovascular Event Associated with Ramp Study Echocardiogram
[摘要] The incidence of ramp test echocardiogram-associated embolic events in the setting of therapeutic anticoagulation is likely rare and has not been reported. We present such a case in a patient with a HeartMate II left ventricular assist device (LVAD) whose serial head computed tomography images, deteriorating clinical course, and the multiembolic nature of the event suggest causality. If the pretest probability of pump thrombosis in an individual LVAD patient is sufficiently high, the potential risks of performing a ramp study echocardiogram may not be warranted, even in the setting of adequate anticoagulation.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:38      统一登录查看全文      激活码登录查看全文