Simple and inexpensive technique for measuring oxygen consumption rate in adherent cultured cells
[摘要] Measurement of cellular oxygen consumption rate (OCR) is essential in assessing roles of mitochondria in physiology and pathophysiology. Classical techniques, in which polarographic oxygen electrode measures the extracellular oxygen concentration in a closed measuring vessel, require isolation and suspension of the cell. Because cell functions depend on the extracellular milieu including the extracellular matrix, isolation of cultured cells prior to the measurement may significantly affect the OCR. More recent techniques utilize optical methods in which oxygen-dependent quenching of fluorophores determines oxygen concentration in the medium at a few microns above the surface of the cultured cells. These techniques allow the OCR measurement in cultured cells adhered to the culture dish. However, this technique requires special equipment such as a fluorescence lifetime microplate reader or specialized integrated system, which are usually quite expensive. Here, we introduce a simple and inexpensive technique for measuring OCR in adherent cultured cells that utilizes conventional fluorescence microscopy and a glassware called a gap cover glass.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 生理学
[关键词] Oxygen consumption ;Mitochondria ;Respiration ;Oxygen measurement  [时效性] 