The Effects of Body Position on Pelvic Tilt Angles Measured with a Smartphone-Based Navigated Ultrasound System in Symptom-Free Young Adults
[摘要] The study aimed to investigate whether pelvic tilt (PT) angles differ between the supine and upright position in symptom-free young adults. Additionally, the concurrent validity of the measurement system was tested on a pelvis phantom against a digital inclinometer. A new smartphone-based navigated ultrasound system was used to perform the measurements. The sample consisted of 12 symptom-free young adults. A multivariate regression model was used to analyze the data. The subjectsâ pelvis in supine position was significantly more tilted anteriorly (mean PT = â7.3°, 95% C.I.: â10.6 to â3.9) than in upright position (mean PT = 0.8°, 95% C.I.: â2.5 to 4.1) (mean. diff. 8.1°; p < 0.001). Rater and trial order had no significant effects on the measured PT angles (p = 0.5). The accuracy of the system when measuring PT angles on the pelvis phantom was 0.3° (0.1° to 0.7°) and 0.2° (â0.09° to 0.6°) for the supine and upright positions respectively. Pelvic tilt angles differed significantly between the supine and the upright position in symptom-free young adults. Concurrent validity showed no differences for measurements in the upright position and small (under 0.4°) significant differences for measurements in the supine position.
[发布日期] [发布机构]
[效力级别] [学科分类] 力学,机械学
[关键词] ultrasonography;computer assisted orthopedic surgery;total hip replacement;anterior pelvic plane;pelvic tilt [时效性]