Evaluation of biomechanical properties of anterior atlantoaxial transarticular locking plate system using three-dimensional finite element analysis
[摘要] PurposeTo evaluate a new anterior atlantoaxial transarticular locking plate system using finite element analysis.MethodsThin-section spiral computed tomography was performed from occiput to C2 region. A finite element model of an unstable atlantoaxial joint, treated with an anterior atlantoaxial transarticular locking plate system, was compared with the simple anterior atlantoaxial transarticular screw system. Flexion, extension, lateral bending, and axial rotation were imposed on the model. Displacement of the atlantoaxial transarticular screw and stress at the screw–bone interface were observed for the two internal fixation systems.ResultsScrew displacement was less using the anterior atlantoaxial transarticular locking plate system compared to simple anterior atlantoaxial transarticular screw fixation under various conditions, and stability increased especially during flexion and extension.ConclusionsThe anterior atlantoaxial transarticular locking plate system not only provided stronger fixation, but also decreased screw-bearing stress and screw–bone interface stress compared to simple anterior atlantoaxial transarticular screw fixation...
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 骨科学
[关键词] Atlantoaxial fusion ;Anterolateral retropharyngeal approach ;Biomechanics ;Finite element analysis  [时效性] 
   浏览次数:1      统一登录查看全文      激活码登录查看全文