An experimental study of ultrasonic imaging with a reduced number of array elements using the envelope method
[摘要] References(23)Cited-By(1)In this experimental study, we conducted measurements using a 128-element ultrasonic array system. Ultrasonic images generated by the conventional migration method were compared with the Envelope method, a promising high-resolution imaging technique. To determine the image quality of an ultrasonic measurement system that uses fewer elements, we selected 64, 32, 16 and 8 elements from the total 128 elements used in our imaging process. The results show that in homogeneous environments the Envelope method can produce clearer images than those produced by the conventional method even when using a small number of elements, suggesting that the introduction of the Envelope method could be an effective way of reducing the number of elements employed in array systems, while maintaining high image quality.
[发布日期] [发布机构]
[效力级别] [学科分类] 声学和超声波
[关键词] Ultrasound CT;Envelope method;Ultrasound;Medical imaging [时效性]