Voltammetric determination of amitriptyline in pharmaceutical formulation with boron-doped diamond electrode exploiting measures in acid medium
[摘要] A simple and reliable voltammetric method is presented for the determination of amitriptyline using a boron-doped diamond electrode in 0.1 mol L1 sulfuric acid solution as the support electrolyte. Under optimized differential pulse voltammetry conditions (modulation time 5 ms, scan rate 70 mV s1, and pulse amplitude 120 mV), the electrode provides linear responses to amitriptyline in the concentration range 1.05 to 92.60 µmol L1 and at a detection limit of 0.52 µmol L1. The proposed method was successfully applied in pharmaceutical formulations, with results similar to those obtained using UVvis spectrophotometric method as reference (at 95% confidence level), as recommended by the Brazilian Pharmacopoeia.
[发布日期] [发布机构]
[效力级别] [学科分类] 化学(综合)
[关键词] differential pulse voltammetry;quality control;BDDE;tricyclic antidepressant. [时效性]